Santosh Kumar Sinha*, Puneet Aggarwal and Anupam Singh
 
Department of Cardiology, LPS Institute of Cardiology, G.S.V.M. Medical College, Kanpur, Uttar Pradesh, India, Email: fionasan@rediffmail.com
 
*Correspondence: Dr. Santosh Kumar Sinha, Department of Cardiology, LPS Institute of Cardiology, G.S.V.M. Medical College, Kanpur, Uttar Pradesh, India, Tel: 91-9670220088, Fax: +91-0512-255619, Email: fionasan@rediffmail.com

Received: 20-Sep-2017 Accepted Date: Dec 13, 2017; Published: 15-Dec-2017, DOI: 10.4172/2368-0512.1000099

Citation: Sinha SK, Aggarwal P, Singh A. Endovascular reconstruction of primary coarctation in a 12-year-old boy technical consideration. Curr Res Cardiol 2017;4(4):73-75.

This open-access article is distributed under the terms of the Creative Commons Attribution Non-Commercial License (CC BY-NC) (http://creativecommons.org/licenses/by-nc/4.0/), which permits reuse, distribution and reproduction of the article, provided that the original work is properly cited and the reuse is restricted to noncommercial purposes. For commercial reuse, contact reprints@pulsus.com

Abstract

Coarctation is a generalized disease which mainly manifests in aorta and presents as aortic coarctation. While transcatheter approaches have replaced surgery, there is still no consensus about the best among only balloon dilatation, aortic stent-grafts, self-expanding stent or balloon expandable stent. Here, we report a case of 12-year-old boy who presented with hypertension in whom severe primary coarctation of thoracic aorta was diagnosed which was successfully treated with endovascular reconstruction using self-expandable stent. Stent migration is always an issue during deployment and which is minimized with rapid pacing and bringing down the blood pressure. We successfully performed the endovascular treatment without aid of pacing thereby bringing his blood pressure to near normal level with no further catch up on follow up. Therefore, balloon expandable sten

Keywords

Thoracic coarctation; Endovascular reconstruction; Self expandable stent; Stent graft; Stent migration

Coarctation of the aorta (CoA) accounts for 5% to 10% of all congenital cardiovascular malformations with, 90% mortality by fifth decade if left unrepaired, mean age being third decade [1]. Various therapeutic options available which include surgery, balloon angioplasty and endovascular reconstruction [2]. The endovascular approach for treating of both native and recurrent coarctation especially among adolescents and adults has become widely acceptable option since the mid-1990s, though optimal approach still remains debatable due to lack of randomized trials between the two. Peripheral balloon angioplasty has the potential to for injury to intima and part of the media with long term potential of recoil, restenosis and aneurysm formation although vessel diameter increases. Therefore, stent implantation scores over balloon angioplasty alone as its radial support to the vessel wall prevents above complication, with the future option of redilatation (especially in young patients), negating the need of over sizing which further avoids any major transmural injury [3,4].

Case Report

A 12-year-old boy presented with history of exertional dyspnoea (NYHA functional class II) of nine months duration. On physical examination, the blood pressure in both arms and both legs were 192/96 mmHg and 92/60 mmHg respectively by appropriately sized cuff. There was absent pulse on tibial and femoral arteries on both legs. Radio-femoral delay was appreciable with a pressure gradient of 100 mm Hg between upper and lower extremities. Other examination findings were not remarkable. Routine haemogram and urine examination were normal. Electrocardiogram revealed left ventricular hypertrophy with mild ST-T changes. 2-D echocardiogram diagnosed a bicuspid aortic valve, left ventricular hypertrophy and normal ejection fraction. He was treated with antihypertensive drugs including amlodipine, losartan, chlorthalidone and prazopressin. A contrast computed tomographic angiogram showed a severe CoA of the descending thoracic aorta at level of the isthmus involving a 3-cm long segment (Figures 1A and 1B).

current-research-cardiology-thoracic

Figure 1: (A) MDCT image of severe thoracic coarctation; (B) Reconstructed image post endovascular reconstruction at 9 months

Catheter based angiogram was done through right radial route which showed severe CoA with profuse collaterals in its vicinity (Figures 2A and 2B).

current-research-cardiology-aortogram

Figure 2: (A, B) Catheter based aortogram through transradial route showing severe thoracic coarctation with multiple collaterals

On further pull back, it showed a gradient of 100 mmHg. Endovascular approach was planned after informed consent. Access was gained by 6F sheath through the right femoral artery (retrograde approach) using micro puncture set and 4000 U of heparin was administered. 6F Judkins right (JR) catheter was advanced over 0.035-inch straight tip Terumo wire (Terumo Inc, Japan) and lesion was crossed (Figure 3A).

current-research-cardiology-judkins

Figure 3: (A) Lesion crossed with judkins right catheter; (B) Stent delivered across the coarcted segment over superstiff Amplatz wire

JR catheter was advanced beyond the lesion and wire was exchanged with 0.035” exchange length (260 cm) superstiff Amplatz guidewire (Meditech, USA). Predilatation was performed by 6 mm × 40 mm Mustang balloon followed by 10 mm × 40 mm Mustang balloon (Boston Scientific, USA).

Pressure gradient trans-aorta after the balloon dilatation was 60 mm Hg. Nitroglycerin infusion was started (15 mcg/mint) to lower down the blood pressure. 12 mm × 60 mm Epic self-expanding stent (Boston scientific, USA) was delivered across the lesion over the wire and gradually released by pulling the lock (Figure 3B). During deployment, wire-stent assembly was held firm as stent was getting migrated. During deployment, stent was also pulled in between to keep it across the lesion and this was performed thrice before final deployment. It was further post dilated with 12 mm × 40 mm Mustang non-compliant balloon at 14 atm pressure (Figures 4A-4D).

current-research-cardiology-dilatation

Figure 4: (A) Well deployed stent showing mild narrowing at the waist; (B) Digital subtraction angiogram (DSA) image showing mild narrowing at the distal edge; (C) Post dilatation of stent with non-compliant balloon; (D) Well apposed and well deployed stent

During catheter pull back, residual gradient was 3 mm Hg. Marked improvement in his symptoms were noted. His blood pressure was 152/80 mmHg with no appreciable gradient between the upper and lower extremities. Femoral and tibial pulses were palpable on both legs. Patient was started on aspirin 75 mg once daily and was discharged with ramipril 5 mg. After 9-months’ follow-up, blood pressure was stable at 136/60 mmHg and patient was asymptomatic and MDCT showed well deployed stent (Figure 1B).

Discussion

The long-term complications of untreated CoA in adults include hypertension, premature coronary artery disease, sudden cardiac death, heart failure, endocarditis, rupture, dissection, aneurysm formation and stroke. Although surgical repair was the gold standard, it has now been replaced by endovascular reconstruction as surgery carries the risk of increased bleeding due to presence of significant intercostal collaterals, paradoxical hypertension, aortic dissection, left recurrent laryngeal nerve paralysis, phrenic nerve injury, and subclavian steal.

Outcomes with balloon angioplasty although shows good initial relief of stenosis but high rate of recoarctation and aneurysm formation therefore, stenting is preferred. Short term efficacy of stent implantation is well recognized, but accurate long term follows up assessment is sparse. Recent reports indicate potential physiological benefits stenting, including midterm left ventricular mass regression and long-axis function improvement, and amelioration in central aortic function with associated reduction of ambulatory systolic blood pressure [5-8].

95% of patients don’t require repeat intervention at 5 years, and short-term complications have been reported in 8.3% of patients among endovascular reconstruction group [9]. The common long-term complications are edge dissection, aneurysm formation, re-coarctation or stent fracture [9-13]. The recurrence of coarctation is attributed to the severity of the disease, fibrotic changes, cystic medial necrosis, intimal tear, thrombosis and graft-infolding in the involved redundant aortic segment due to over sizing during stent deployment. Therefore, stent should not be oversized as they remodel in long term and the purpose should be to abolish the gradient. Most of the available data suggest use of covered stent but it is very costly and not readily available, therefore we decided to use self-expanding stent [4]. However, certain substrate warrants covered stents such as: subatretic coarctation, native coarctation associated with patent ductus arteriosus, re-coarctation combined with aneurysm formation [14]. Sometimes, right ventricular pacing at fast rate (or rarely adenosine infusion) is employed during inflation to facilitate the precise stent placement [13]. We used only nitroglycerin drip to lower down the blood pressure and deployed the stent without other aids. As stents remodel with time, post dilatation at high pressure using non-compliant balloon is not required except if a waist or any significant pressure gradient persists [12]. Also, one should avoid flaring the ends of the stent with a larger balloon to attain geographical match with the dilated post-stenotic segment because there is no clear-cut evidence of any significant benefit, but it carries a real risk of stent migration or aneurysm formation as shown by Chessa et al. [12]. The procedural success is defined as residual gradient <10 mmHg, and an improvement in diameter to 90% of the normal adjacent aortic arch vessel [13].

Conclusion

The diameter and length of stent are most important factor for technical consideration as the stent must cover the affected segment, and at the same time proximal end not protruding into the left subclavian artery. The stent diameter must not exceed the diameter of the descending aorta at the level of the diaphragm which is considered as reference level in order to avoid aneurysm formation or rupture [4]. Alternatively, the average diameter of the aortic isthmus (or transverse aortic arch) and the descending aorta at the level of diaphragm may also be considered. Also, the ratio of expanded stent diameter to that of coarcted segment is usually kept around 2-3.5 and should always be <3.5 to avoid aneurysm formation or dissection as stents remodel with time. The most important parameter should be relieving of pressure gradient. Therefore, endovascular reconstruction for thoracic coarctation with self-expanding stent is a safe and promising with acceptable alternative to surgical repair or balloon angioplasty alone.

Acknowledgements

Vinay Singh-Radiation Safety Officer for his technical support in MDCT image reconstruction.